Configure an Experiment Method for Concentration
The absorbance of a solution is related to the concentration of the species within it. The relationship, known
as Beer’s Law , is:
A
λ
=
ε
λ
c
l
where
A
is the absorbance at wavelength
λ
,
ε
is the extinction coefficient of the absorbing species
at wavelength
λ
, c is the concentration and
l
is the optical pathlength.
Concentration is the amount of a specified substance in a solution. Graphs of absorbance vs. concentration
are known as Beer’s Law plots or concentration calibration curves. These are prepared by measuring the
light absorbed by a series of solutions with different known concentrations. The length of the sample --
such as the path length of your cuvette -- and the wavelength chosen for monitoring the amount of light
absorbed are constants. A plot from taking scans of these standard solutions with known concentrations is
then obtained. This plot is used to determine the unknown concentrations of substances in solutions.
"
""
"
Before you begin, we suggest you place a blank in the sampling region, store a reference and a
dark, and then place your sample with the highest known concentration in the sample holder of
the Palm-SPEC. Make sure you are in absorbance mode. Find the wavelength of the highest
peak. Note this wavelength.
1. Select
Method | Create
from the menu. You receive a
Create Method
dialog box where you must
name the method you are about to create. After naming the file, click
OK
. The method you now create
is saved under this file name.
2. An empty
Edit Method
dialog box opens. You now need to select and insert the software events you
need for your experiment method. Select the
Insert
button in the right side of the dialog box to view
all 104 possible software events.
3. The
Edit Method Step
dialog box opens. Under
Method Action
, click on the down arrow to reveal
104 software events. You have to insert one event at a time. The events recommended to insert for a
concentration experiment and the order that they should appear are the following:
0: Display cursor value on main screen
This step is optional. Selecting this step allows you to see the absorbance values of your solutions with different
known concentrations, which helps to verify that the experiment is working correctly.
1: Set cursor location
This step allows you put the cursor at a specific wavelength for the absorbance measurements of your
samples with known concentrations. One option is to place the cursor on the wavelength noted in the
"
""
"
message above. Another option is to select a wavelength that you know is in a non-absorbing region of
the samples.
2: Set acquisition parameters
This step is also optional. However, if you do not set acquisition
parameters for your method, then the software will use the acquisition
parameters already in use, which are set by selecting
Spectra | Configure
| Data Acquisition
from the menu. For details on selecting values for
integration time, averages, and boxcar width, see page 17.
3: Store reference and dark spectra
You must select this step as storing a reference and a dark is requisite
before the software can calculate absorbance. This command merely
stores a reference and a dark. If you want to save the reference and dark,
insert the
Save reference spectrum
and
Save dark spectrum
method steps.
4: View Data in absorbance mode
This step is mandatory. It assures that you are in absorbance mode while the Palm-SPEC analyzes your
solutions with known concentrations.