5. When you are ready to execute the concentration method, make sure that you have a blank in the
sampling region but that the series of solutions with different known concentrations are at hand, as you
will receive prompts to insert these samples into the cuvette holder.
6. Now select
Method | Execute
from the menu. A dialog box containing method files opens. Select the
correct concentration method and click
OK
.
7. As the Palm-SPEC reads the method steps and executes each step in your
concentration method, you will notice the light source turn off (as it
executes method step
3: Store reference and dark spectra
) and the view
switches to Absorbance mode (as it executes method step 4:
View data in
absorbance mode
). You then receive a
Calibration Point
message
prompt similar to the one at right. Make sure that you place the solution
with the correct concentration value into the sample chamber of the Palm-SPEC. Then click the
OK
button in the dialog box.
8. After the Palm-SPEC scans the first sample, you receive a
Calibration Point Stored
message verifying that the sample was
scanned and stored. Click
OK
.
9. You will continue to receive
Calibration Point
message prompts
and
Calibration Point Stored
confirmation messages for each
sample you stored in the concentration method.
10. Once all of your samples with solution of known concentrations have
been scanned, the Palm-SPEC performs the regression. If you inserted
the method step
6: Display regression results
, you receive a
Calibration Complete
message with the intercept, first coefficient,
second coefficient and R
2
values. The R
2
value should be very close to
1.0. If it is not, you may want to perform the method again. Click
OK
.
11. The view mode switches to concentration. You are now ready to insert samples with solutions of
unknown concentrations. When you need to exit out of the concentration mode, select
Method | Exit
from the menu.